LabDry is a system that supports the gravimetric determination of water content by oven drying - the measurement according to the standard, in which the sample is weighed before and after drying. It consists of calculation software for a PC, a handheld barcode scanner and aluminium dishes marked with codes.
The point of the arrangement is a single one: to take human error out of the procedure. The barcode on the dish, a balance connected to the computer and software that does the arithmetic make it impossible to mix up samples, to copy a weight incorrectly or to miscalculate. The result goes straight into the data archive.
The balance and the drying oven are not part of the set and are chosen separately. LabDry itself neither weighs nor dries - it supplies the equipment and the logic that tie both operations into a documented measurement. Both groups of devices are described below.
Scope of delivery
Art. no. 12176 - LabDry, system for the determination of water content. |
![]() |
All three ovens work with forced air circulation and serve the determination of water content by oven drying according to the standard. What separates them is the capacity, that is the number of samples dried at one time.
| Art. no. | Oven | Usable volume | Number of samples | Internal dimensions | External dimensions | Weight |
|---|---|---|---|---|---|---|
| 12324 | Drying oven 56 | 60 l | up to 12 at a time, up to 20 with two additional trays | 400 × 345 × 440 mm | 560 × 565 × 625 mm | 39 kg |
| 12322 | Drying oven 115 | 116 l | up to 21 at a time, up to 42 with three additional trays | 550 × 385 × 550 mm | 710 × 605 × 735 mm | 50 kg |
| 12328 | Drying oven 260 | 259 l | up to 30 at a time, up to 90 with six additional trays | 650 × 515 × 780 mm | 810 × 760 × 965 mm | 85 kg |
A chrome-plated slide-in shelf is ordered separately for each oven: 13716 for oven 56, 12913 for oven 115, 13830 for oven 260.
The balance must have an interface to the computer - without it the weight does not reach the software automatically and the whole point of the arrangement is lost. All three models come with one, cable included.
| Art. no. | Balance | Weighing range | Readability | Repeatability | Linearity | Notes |
|---|---|---|---|---|---|---|
| 15073 | Precision balance 2000 | 2000 g | 0.01 g | 0.01 g | ±0.03 g | weighing pan 128 × 128 mm |
| 15074 | Precision balance 3000 | 3100 g | 0.1 g | 0.1 g | ±0.3 g | weighing pan 128 × 128 mm |
| 12499 | Precision balance 3500 | 3500 g | 0.01 g | 0.01 g | ±0.02 g | approved for verification and initially verified, alibi memory for 100 000 results, 230 V / 50 Hz supply |
The choice comes down to two questions: whether the measurement has to be documented with a verified balance (then 12499) and whether a readability of 0.1 g is enough (then the cheaper 15074).